Ax60+ CO₂ Calibration Guide

Proof Testing and Calibration Explained

Regular calibration of CO₂ gas sensors is essential to ensure accurate readings and maintain a safe working environment. If gas monitors are not correctly maintained, they may fail to detect dangerous conditions, putting people and operations at risk.

This guide explains how to proof test and calibrate your Analox Ax60+ CO₂ sensor, helping you maintain accuracy, compliance, and safety.

What is Gas Sensor Calibration?

Calibration is the process of comparing a gas sensor’s readings against a known reference gas concentration to ensure accuracy.

In simple terms, calibration ensures your gas monitor is:

  • Measuring correctly
  • Responding to gas exposure
  • Providing reliable safety alerts

Why is Calibration Important?

Without proper calibration, gas sensors can drift over time. This can lead to:

  • Inaccurate readings
  • False alarms
  • Failure to detect dangerous gas levels

Regular calibration helps:

  • Maintain accuracy
  • Reduce downtime
  • Protect workers
  • Demonstrate compliance with safety standards
Graph showing improved gas monitor accuracy after calibration of CO2 sensor

Proof Test vs Calibration: What’s the Difference?

Understanding the difference is key:

Proof Test (Bump Test)

A quick check to confirm the sensor responds correctly to gas.

Calibration (Zero & Span)

Adjusts the sensor to restore accurate readings.

If a proof test fails, calibration is required.

How Often Should You Calibrate?

We recommend:

  • Proof test: At least every 12 months
  • Calibration: Only if the proof test fails or drift is detected

You should also calibrate if:

  • A high alarm has occurred
  • Incorrect gas was used
  • Maintenance has been neglected

Equipment Required

Before starting, ensure you have:

  • Push-in flow adaptor
  • 2m neoprene tubing
  • 0.5–1 L/min regulator
  • Calibration gas (3.0% CO₂, balance nitrogen)
  • Zero gas (20.9% oxygen or clean air)

Analox offers a complete Ax60+ CO₂ Check Kit with all required components.

Melting ice in ocean illustrating impact of CO2 emissions and climate change

How to Perform a Proof Test (Bump Test)

  1. Ensure the Ax60+ is powered and operating normally
  2. Attach tubing and flow adaptor to the sensor
  3. Connect regulator to calibration gas
  4. Apply gas for 1–2 minutes
  5. Check reading is 30,000–30,150 ppm CO₂
  6. If within range - no action required
  7. If not - proceed to calibration

Zero Calibration (First Step)

  1. Access calibration menu (hold Cancel + Cycle)
  2. Select Zero calibration
  3. Connect zero gas or use clean air
  4. Apply gas for 1–2 minutes
  5. Wait for reading to stabilise
  6. Confirm calibration

Always perform zero calibration before span calibration.

Analox Ax60 CO2 monitor central display showing calibration menu navigation using control buttons

Span Calibration

  1. Select Span calibration in the menu
  2. Set value to 3.0% CO₂
  3. Apply span gas for 1–2 minutes
  4. Wait for reading to stabilise
  5. Confirm calibration

Troubleshooting Tips

  • Only calibrate one sensor at a time
  • Ensure gas flow is not blocked
  • Remove any splash guards before testing
  • If calibration fails, check fault codes in the manual
Analox Ax60 CO2 sensor with removable splash guard for gas monitoring applications

When Should You Call an Engineer?

If you don’t have the correct equipment or are unsure, your gas supplier or Analox can provide:

  • On-site calibration
  • Service support
  • Replacement components

FAQs About Ax60+ Calibration

What is a bump test?

A quick test using calibration gas to confirm the sensor responds correctly.

How often should I calibrate my Ax60+?

Annually, or whenever a proof test indicates drift.

Is there a video to show me how to calibrate the Ax60+?

Yes, please see below:

What happens if calibration fails?

A fault code may appear, indicating the need for further checks or servicing.

Can I calibrate the Ax60+ myself?

Yes, the system is designed to be user-friendly for operators and engineers.

Why is calibration important?

It ensures accurate readings, protects workers, and prevents dangerous gas exposure.

To learn more about the Ax60+, please click below:

Key takeaway

Maintaining your Ax60+ CO₂ sensor through regular proof testing and calibration is essential for ensuring accurate gas detection.

A properly calibrated system helps:

  • Protect workers
  • Maintain compliance
  • Prevent costly incidents

If your environment uses CO₂, regular calibration is not optional, it’s critical for safety.