Aspida

Personal CO₂/O₂ Monitor

The Aspida is the ideal solution to protect you and your team from the dangers of enriched oxygen (O₂), carbon dioxide (CO₂) or inert gas leaks.

The Aspida is a robust, high specification personal oxygen (O₂), carbon dioxide (CO₂) or dual (O₂ / CO₂ combined) monitor which can be worn on a belt or even wall mounted as a backup to a primary safety system.

The Aspida, (Greek for Shield) helps you comply with the regulations, will remind you when maintenance is needed and it safeguards the user from the gas risks.

When you need hassle free, personal protection, choose the Aspida.

Key Features

  • IP65 rated, ensuring the unit is dust tight and protected against water
  • Data logging capability
  • Man down alarm
  • Time Weighted Average (TWA) monitoring
  • Rechargeable batteries
  • Intelligent software – up to 2 users registered to one unit
  • User changeable alarms
  • We recommend: 23.5% for oxygen (O₂) enrichment (As per NHS SOPS/EH40/OSHA); 19.5% for oxygen (O₂) deficiency 0.5%; & 1.5% for carbon dioxide (CO₂) enrichment

Information

Datasheets & Manuals

Legislations / Standards

Standards such as OSHA, EH40 and TUV in Europe mandate that employees and members of the public are not exposed to potentially dangerous levels of CO₂ as it is a highly toxic gas in relatively small quantities.

Guidelines also advise consideration be given to the use of oxygen monitors where inert gases such as nitrogen are used, as a leak can cause an asphyxiation risk.

The NHS SOPS (Standard Operating Procedures) also details the requirement for oxygen (O₂) enrichment monitoring.

Analox are committed to helping you achieve your local and application specific legislation, please call us to discuss your requirements.

Technical Specifications
CO2 SensorO2 Sensor
Sensor TypeAnalox infrared MIRElectrochemical
Range0.01 to 5%0.1 to 25%
Accuracy (at standard temperature and pressure)±2% of full scale between 0.01% and 2.5%, ±3%±3% of full scale between 0.1% and 25%
Response timeT90 <60 secondsT90 <30 seconds
Sensor life span5 years2 year (expected)

Additional Information

Personal Protection

Enriched oxygen (O₂ can cause a condition on the human body called hyperoxia, enriched CO₂ is highly toxic, both are a threat to life and must be safeguarded against.
If an inert gas such as nitrogen (N₂), argon (Ar), helium (He) or liquid nitrogen (LN2) leaks then this will deplete the levels of oxygen (O₂) in the surrounding area, creating a dangerous atmosphere.

The Aspida can offer personal protection and alerts to all of these risks, offering you true peace of mind.

Ideal Solution

Offering audible and visual alarms, data logging and a man down alarm for lone workers, it is an ideal solution to protect staff from the dangers of a leak of carbon dioxide (CO₂), oxygen (O₂) or inert gas.

The Aspida is an affordable, easy to operate gas detection monitor and is perfect for ensuring safety in areas where the gas is piped, used or stored.

It operates using rechargeable battery technology allowing it to run for more than 12 hours continuously between charges.

Applications
  • Breweries
  • Restaurants
  • Hospitality
  • Laboratories
  • Medical
  • Agriculture
  • Oil & Gas
  • Transportation

Obsolete Product Manuals

Frequently Asked Questions

How do I proof test my Aspida oxygen sensor?

The process for confidence testing the Aspida units is similar to the calibration process as described in the product user manual.

The procedure below is similar to testing carried out by Analox when validating the devices are operating within specification:

  1. Connect the supplied flow adaptor to the device (PGA-401DK) or connect the free end of the flow adaptor hose to a flow regulator (SA7L705XFLIND)
  2. Fit the flow regulator to a cylinder of 20.9% O₂ (In N₂) (SA7L11009)
  3. Set the flow rate to 0.2lpm
  4. Leave to settle for 3 – 5 minutes
  5. Carry out a calibration of the unit as described in the product user manual
  6. Ensure the oxygen reading is within ±0.75% of 20.9% O2
  7. Change the gas cylinder to a 100% N₂ cylinder (SA7L10308)
  8. Set the flow rate to 0.2lpm
  9. Leave to settle for 3 – 5 minutes
  10. Verify that the O₂ reading begins to fall and at about 19.5% the following activate: (Red LEDs, Sounder, Vibration motor (stops vibrating after 10 seconds), O₂ reading shows a flashing outline and flashing symbol.
    Note. 19.5% is a default O₂ alarm level, this may vary if your device settings have been changed
  11. Verify that at 18% O₂, the LEDs and sounder begin to pulse faster
    Note. 18% is a default O₂ alarm level, this may vary if your device settings have been changed
  12. Press the unnamed 5 button to mute the alarm
  13. Note the O₂ reading is correct to ±0.75% of 0% O2
How to Calibrate the Analox Aspida