Ax60+

Modular Gas Safety System

The Ax60+ is a time served, trusted gas safety, modular gas safety system capable of measuring CO₂, O₂ and/or CO, and can be used across many different industries. You can build the system to your requirements as the bespoke modular setup provides the flexibility to integrate up to four different sensors (per setup) for the simultaneous measurement of various gases.

There is a common misconception that if using carbon dioxide (CO₂) in your facility you can stay safe by using an oxygen (O₂) monitor. The team here at Analox want to dispel this myth so we can ensure you and your team stay safe when using and storing CO₂.

CO₂ is a toxic gas which, even at relatively low levels, can have dangerous effects on the human body. In the event of a CO₂ leak, the CO₂ will begin to displace the O₂ in the area, however, O₂ sensors typically have alarm levels set at around 19.5% and 18% so by the time the CO₂ has displaced enough oxygen to trigger these alarm levels the CO₂ levels will be dangerously high.

If using CO₂, your safety monitor must be a CO₂ monitor, this is critical so you are alerted to any leaks at levels which still give you time to take action to keep everyone safe.  This also helps to save you money as leaks can be addressed early and in the current times of CO₂ shortages, protecting this precious resource is key to any business.

Information

Installation

How should you install the Ax60+ unit for optimal safety monitoring, and where is the best location within your facility?

Technical Specifications
Ax60+
FeatureSpecification
Warm-up Time40 seconds
Operating Temperature Range5°C to +50°C
+23°F to +122°F
Warranty5 Years Electronics
DisplayLCD dot matrix (backlit)
Relays2 x high powered configurable relays rated for 250V AC/ 30V DC, 3A max
Maximum Power Consumption<24 W
Power Supply Accepts Inputs From100-240V ~ 1.5A through the transformer or 24VDC direct power
IP RatingIP54
DimensionsW: 6.9” / 175mm
H: 4” / 106mm
D: 2.6” / 67mm
Optional Extras24V supply available for optional external beacon
Alarm
FeatureSpecification
StrobeWhite LED strobe > 100 cd (optional blue, red & amber filters)
SounderMin 80 dBA @3m / (9.8 feet)
Warranty5 Years Electronics
DimensionsW: 3” / 72mm
H: 5.1” / 132mm
D: 2” / 52mm

Additional Information

Ax60+ Central Display

The Central Display Unit for the Ax60+ is required for all builds (except when using an Ax60+ Kiosk which is a standalone sensor).

The unit has been specifically engineered for use in environments where accurate gas analysis is critical. The Ax60+ CDU clearly displays the information you need and helps to ensure the safety of personnel by continuously monitoring the levels of various gases.

The Ax60+ features an intuitive and user-friendly interface, making it easy for operators to navigate and access critical information. The clear display provides real-time data on gas concentrations, allowing for quick and informed decision-making.

This unit acts as a main panel for up to 4 sensors. The LCD (Liquid crystal display) screen is used to display the readings for each sensor as well as any fault codes that may appear when the sensor is in fault. Also within the central display unit you can attach a beacon, a DOM (Data Output Module) unit and up to 2 relays on one system.

From the Central Display Unit, the user can access the technical menu, this is where you can customise your Ax60+ to your own needs, and also from these same menus you can calibrate your sensors to make it more user maintainable. The Ax60+ also comes with a universal power supply, making it adaptable to any situation.

Ax60+ Alarm Units

The Ax60+ Alarm Units are integral components of a comprehensive gas detection system, working together with the Central Display and Sensor units to enhance safety in environments where accurate gas monitoring is crucial. These alarm units are specifically designed to provide immediate and unmistakable alerts in the event of abnormal gas concentrations, ensuring a swift response to potential hazards.

The Ax60+ Alarm Units are equipped with both audible alarms and highly visible LED indicators. This dual-alert system ensures that operators are promptly notified of any deviations from acceptable gas levels, even in noisy or busy industrial settings.

Applications
  • Breweries & Wineries
  • Hospitality
  • Laboratories
  • Medical
  • Oil & Gas
  • Construction
  • Indoor Spaces
  • Indoor Farming
  • Cultivation

Obsolete Product Manuals

Frequently Asked Questions

How to silence and acknowledge latched alarms on the Analox Ax60+

The Ax60+ systems have latched alarms as standard however does have the option of disabling alarm latching. It is our hope that the following information will help to avoid confusion when silencing and acknowledging alarms.

Latched Alarms

These will not clear unless acknowledged. For example, if the target gas level increases beyond the alarm threshold, the device will go into alarm and remain like this until the alarm has been acknowledged.

Non-Latched Alarms

These will clear without acknowledgement. For example, if the target gas level increases beyond the alarm threshold, the unit will go into alarm but once the gas returns to safe limits the alarm automatically clears.

Both Latched and Non-Latched alarms can be muted with a single press of the Accept/Test button, but in order to acknowledge a latched alarm the user must first cycle through the display until an ‘*’ is seen, then press and hold the Accept/Test button for 3 seconds (resulting in the ‘*’ changing into a ‘>’.

The video below shows an example of muting and acknowledging a latched alarm:

Can the Ax60+ Alarm Unit be mounted outside, and is it weather tight?

The Ax60+ Alarm/Strobe Unit is IP55 rated.

IP means Ingress Protection, the 2 numbers after indicate to what level the unit is able to protect itself against dust, moisture, liquids and accidental contact. The IP rating for this unit is broken down below;

  • 5 – Solids – Dust Protected. Limited ingress of dust permitted. Will not interfere with operation of the equipment. 2 to 8 hours.
  • 5 – Water – Protected against jets of water. Limited ingress permitted.

However, the unit is Not Waterproof/Weatherproof for long periods of time, and in direct sunlight the UV light will fade the label on the unit. The Alarm Unit can be mounted outside if it has sufficient protection from the elements.

How do I disable the Alarm sounders on the Ax60+?

There are two potential reasons for the Ax60+ to show a FLT16 code on the screen, these are:

  1. The sensor unit requires zero and span calibration adjustment (see the service manual for guidance on carrying out this process)
  2. The CO₂ sensor alarm directions have been incorrectly set.

If your unit is displaying a Flt16, first attempt a zero and span calibration adjustment then power cycle the device. If this is not possible or the issue is not resolved, check to see if it has been correctly configured as detailed below.

The Ax60+ is primarily a breathing air device, as such, the CO₂ sensor unit alarms are restricted to only trigger in the ‘high-going’ direction. If a user changes the alarm ‘going’ to ‘low-going’ a FLT16 will ensue.

To clear this fault the following steps should be followed (be aware that only a short period of non-alarm time is available so you’ll need to be quick!!):

  1. Power cycle the Ax60+ (turn it off and on again)
  2. Wait for the sensors display on the screen
  3. Press the Accept/Test button briefly to silence any alarms. Press and hold the top two buttons (Cancel and cycle) until the top level menu is reached.
  4. Press the cycle button six times (until sensor config is shown)
  5. Press the Accept/Test button. Press the cycle button twice (Alarm direction is shown)Press the Accept/Test button. Select the sensor that had the alarm ‘going’
  6. changed (press cycle to move the arrow selector over until the relevant sensor is identified and press the Accept/. test button)
  7. Select the alarm that had the alarm ‘going’ changed
  8. (press cycle to move the arrow selector over until the relevant alarm is identified and press the Accept/test button)
  9. Change the alarm direction to ‘up’ (press the cycle button until the arrow selector points at the up arrow and press the Accept/test button)
  10. Press the cancel button five times until the sensors display on screen
  11. Acknowledge any faults displayed
  12. Power cycle the unit
How do I fix a loose RJ45 connector on my Analox Ax60+?

What should I do if my Ax60+ has a FLT16 on the display?

There are two potential reasons for the Ax60+ to show a FLT16 code on the screen, these are:

  1. The sensor unit requires zero and span calibration adjustment (see the service manual for guidance on carrying out this process)
  2. The CO2 sensor alarm directions have been incorrectly set.

If your unit is displaying a Flt16, first attempt a zero and span calibration adjustment then power cycle the device. If this is not possible or the issue is not resolved, check to see if it has been correctly configured as detailed below.

The Ax60+ is primarily a breathing air device, as such, the CO2 sensor unit alarms are restricted to only trigger in the ‘high-going’ direction. If a user changes the alarm ‘going’ to ‘low-going’ a FLT16 will ensue.

To clear this fault the following steps should be followed (be aware that only a short period of non-alarm time is available so you’ll need to be quick!!):

  1. Power cycle the Ax60+ (turn it off and on again)
  2. Wait for the sensors display on the screen
  3. Press the Accept/Test button briefly to silence any alarms. Press and hold the top two buttons (Cancel and cycle) until the top level menu is reached.
  4. Press the cycle button six times (until sensor config is shown)
  5. Press the Accept/Test button. Press the cycle button twice (Alarm direction is shown)Press the Accept/Test button. Select the sensor that had the alarm ‘going’
  6. changed (press cycle to move the arrow selector over until the relevant sensor is identified and press the Accept/. test button)
  7. Select the alarm that had the alarm ‘going’ changed
  8. (press cycle to move the arrow selector over until the relevant alarm is identified and press the Accept/test button)
  9. Change the alarm direction to ‘up’ (press the cycle button until the arrow selector points at the up arrow and press the Accept/test button)
  10. Press the cancel button five times until the sensors display on screen
  11. Acknowledge any faults displayed
  12. Power cycle the unit

Is the Ax60+ backwards compatible with the Ax60?

There is some backwards compatibility:

  • Ax60+ CO₂ sensor units CAN be used with both Ax60 and Ax60+ central displays.
  • Ax60 sensor units CAN ONLY be used with Ax60 central displays.
  • Oxygen sensors CAN ONLY be used on the Ax60+ central displays.

See table below for reference:

 

Ax60 Central Display

Ax60+ Central Display

Ax60 CO₂ Sensor

✔

✗

Ax60+ CO₂ Sensor

✔

✔

Ax60+ O₂ Sensor

✗

✔

What do I do if my Ax60+ Unit is in Alarm?

Stay out of the area, open all windows and doors to ventilate. Do not re-enter the area until the alarm condition stops and the reading on the display has returned to normal.

Have a suitably qualified person check the gas bottles/lines for leaks/frosting.

To silence the alarm, press the Accept/Test button.

Ax60+ Comms Fault

A COMMS fault means a sensor has developed a communications fault. May be caused by accidental cable disconnection. To clear a COMMS FAULT press Accept/Test to silence the buzzer then reconnect the cable.

A COMMS fault can also be caused by incorrect jumper positions. Each sensor must have a unique jumper position. Two sensors with the same jumper position will cause a COMMS fault.

  • Sensor 1 must have the jumper in position 1
  • Sensor 2 must have the jumper in position 2
  • Sensor 3 must have the jumper in position 3
  • Sensor 4 must have the jumper in position 4

Ax60+ Modular Gas Safety System

My Ax60+ is displaying FLT06

FLT06 is displayed when the Ax60+ fails to initialise correctly. This often means that the sensor did warm up correctly and did not reach its operating temperature. This can happen when the environment around the Ax60+ is too cold.

Turning the device off and on again should clear the FLT06.

If your device has latching faults enabled, the fault will need to be acknowledged first, to acknowledge the fault, please see the above instructions for acknowledging latched alarms.

My Ax60+ sensor unit got wet and won’t stop alarming

The Ax60+ sensor units are IP55 rated, which means they are water-resistant, but not waterproof.

If the Ax60+ sensor unit gets soaked, it can sometimes cause it to go into alarm in error.

Powering down the sensor unit and giving it time to dry out should solve the problem and the Ax60+ should return to normal operation.

What height should my Ax60+ sensors be installed?
  • CO₂ – 12-18″ above ground
  • Cultivate  – 12-18″ above ground
  • SAFE-OX – Average Head height
  • CO – Average Head height
Why can’t I find the Ax60+ O₂ sensor?

This has now been made obsolete and has been replaced with the Ax60+ SAFE-OX.