Ax60+ DOM

Data Output Module

The Ax60+ Data Output Module, the perfect accessory to help you provide real time monitoring of your Ax60+ system.

The Data Output Module (DOM), designed specifically for the Analox Ax60+, acts as a gateway between the Ax60+ and your remote systems.

The DOM allows the Ax60+ to communicate real time monitoring to platforms such as building management systems (BMS), providing crucial opportunities for in depth trend analysis.

Key Features

    • A 4-20mA signal for each connected sensor (up to four)
    • Digital RS485 MODBUS RTU Communications
    • Simple installation & setup and it is easily retrofitted to existing Ax60+ installations
    • Up to 16 Ax60+ DOMS can be digitally networked supporting up to 64 sensors
    • Number of analogue outputs: 4
    • Output load capability: 500 ohms max loop resistance per output
    • Analogue output type: active 4-20mA per output
    • Digital output: RS-485
    • Comms protocol: Modbus RTU
    • Mounting: wall mounted as per sensor
    • Certification: CE & IEC 61010-1 (UL) plus Australian AS5034-2005
    • Warranty: 2 years

Information

Datasheets & Manuals

Installation

How should you install the Ax60+ unit for optimal safety monitoring, and where is the best location within your facility?

Technical Specifications
FeatureSpecification
Number of analogue outputs4
Output load capability500 ohms max loop resistance per output
Analogue output typeActive 4-20mA per output
Digital outputRS-485
Comms protocolModbus RTU
MountingWall mounted as per sensor
CertificationCE & IEC 61010-1 (UL) plus Australian AS5034-2005
Warranty2 Years
DimensionsW: 3” / 72mm
H: 5.1” / 132mm
D: 2” / 52mm

Additional Information

Retrofit

The DOM can be retro fitted to any existing Ax60+ installations.

The Data Output Module can be used to interface an Ax60+ system into your BMS, providing live readings via RS485 MODBUS RTU or up to 4 independent 4-20mA current loop signals.

Because the DOM provides industry standard analog and digital outputs, many systems (with features such as data logging) will be compatible, enabling you to conduct after-incident reviews or trend analysis. Additionally, these systems can give real-time indication of any readings or alarm conditions measured by the Ax60+.

Building Management Systems

Building Management Systems are integral to the control of the building’s environment. The BMS can automatically take action, based on sensor inputs, eliminating the risk of human error. For example, in the case of a Gas Leak, the system would ventilate the area, isolate gas supply, close interleading doors and notify the necessary authorities, all of this, without human intervention or wasted time.

Without the DOM providing the link from the Ax60+ to the BMS, the building management system could unknowingly move contaminated air into inhabited areas, or automatically turn off ventilation in rooms with high levels of contaminated air.

The BMS is usually the first port of call for all building environmental management. Integrating the Ax60+ System to your BMS reduces the risk to health, increases emergency response time, and can inform the treatment of casualties.

4-20mA

The DOM converts the Ax60+ gas sensor readings into industry standard analog signals which are suitable for interfacing with many commercially available systems.

When the analog levels are 4mA, the equivalent gas level is classed as 0% of the sensor’s range. At 20mA the equivalent gas level is classed as 100% of the sensor’s full scale. Equivalent gas levels can be calculated by taking measured analog signals and scaling them as a proportion of their full scale readings. The DOM can therefore be used to communicate gas levels measured by the Ax60+ sensor units to external systems.

RS485 MODBUS RTU

This industry-standard interface can be used with digital building management systems and allows for advanced and more detailed information on the operational state of the Ax60+ system.

The system status information the DOM makes available includes;

  • Whether sensors are in alarm
  • What level the alarms are at (TWA, high and high high in the case of CO₂)
  • The system’s operational status
  • The gas level detected by each of the sensors.

Larger Installs

For larger installations or sites with multiple Ax60+ systems up to 16 data output modules can be networked together and connected to the building management system meaning a single channel on a BMS can be used to monitor up to 64 sensor units!

Both 4-20 mA outputs and Modbus RTU interface can easily and quickly be connected to a compatible device or system, providing a remote visual indication of measured gas levels. The DOM is a self contained unit that can be connected through CAT 5 cable.

Once integrated into your system, the DOM’s digital communications allow for remote monitoring and in depth trend analysis of monitored gasses.

  • Transmits gas levels (e.g. 5000ppm CO₂)
  • Transmits alarm status (e.g. AL2 Triggered)
  • Transmits fault status (e.g Sensor 2 is in fault)

Applications
  • Beverage
  • Hospitality
  • Laboratories
  • Construction

Obsolete Product Manuals

Frequently Asked Questions

Can the AX60+ Alarm Unit be mounted outside, and is it weather tight?

The Ax60+ Alarm/Strobe Unit is IP55 rated.

IP means Ingress Protection, the 2 numbers after indicate to what level the unit is able to protect itself against dust, moisture, liquids and accidental contact. The IP rating for this unit is broken down below;

  • 5 – Solids – Dust Protected. Limited ingress of dust permitted. Will not interfere with operation of the equipment. 2 to 8 hours.
  • 5 – Water – Protected against jets of water. Limited ingress permitted.

However, the unit is Not Waterproof/Weatherproof for long periods of time, and in direct sunlight the UV light will fade the label on the unit. The Alarm Unit can be mounted outside if it has sufficient protection from the elements.

How do I disable the Alarm sounders?

All relevant Ax60+ document information can be found on the Ax60+ product page.

There are two potential reasons for the Ax60+ to show a FLT16 code on the screen, these are:

  1. The sensor unit requires zero and span calibration adjustment (see the service manual for guidance on carrying out this process)
  2. The CO₂ sensor alarm directions have been incorrectly set.

If your unit is displaying a Flt16, first attempt a zero and span calibration adjustment then power cycle the device. If this is not possible or the issue is not resolved, check to see if it has been correctly configured as detailed below.

The Ax60+ is primarily a breathing air device, as such, the CO₂ sensor unit alarms are restricted to only trigger in the ‘high-going’ direction. If a user changes the alarm ‘going’ to ‘low-going’ a FLT16 will ensue.

To clear this fault the following steps should be followed (be aware that only a short period of non-alarm time is available so you’ll need to be quick!!):

  1. Power cycle the Ax60+ (turn it off and on again)
  2. Wait for the sensors display on the screen
  3. Press the Accept/Test button briefly to silence any alarms. Press and hold the top two buttons (Cancel and cycle) until the top level menu is reached.
  4. Press the cycle button six times (until sensor config is shown)
  5. Press the Accept/Test button. Press the cycle button twice (Alarm direction is shown)Press the Accept/Test button. Select the sensor that had the alarm ‘going’
  6. changed (press cycle to move the arrow selector over until the relevant sensor is identified and press the Accept/. test button)
  7. Select the alarm that had the alarm ‘going’ changed
  8. (press cycle to move the arrow selector over until the relevant alarm is identified and press the Accept/test button)
  9. Change the alarm direction to ‘up’ (press the cycle button until the arrow selector points at the up arrow and press the Accept/test button)
  10. Press the cancel button five times until the sensors display on screen
  11. Acknowledge any faults displayed
  12. Power cycle the unit
What do I do if my Ax60+ Unit is in Alarm?

Stay out of the area, open all windows and doors to ventilate. Do not re-enter the area until the alarm condition stops and the reading on the display has returned to normal.

Have a suitably qualified person check the gas bottles/lines for leaks/frosting.

To silence the alarm, press the Accept/Test button.

My Ax60+ is displaying FLT06

FLT06 is displayed when the Ax60+ fails to initialise correctly. This often means that the sensor did warm up correctly and did not reach its operating temperature. This can happen when the environment around the Ax60+ is too cold.

Turning the device off and on again should clear the FLT06.

If your device has latching faults enabled, the fault will need to be acknowledged first, to acknowledge the fault, please see the above instructions for acknowledging latched alarms.

My Ax60+ sensor unit got wet and won’t stop alarming

The Ax60+ sensor units are IP55 rated, which means they are water-resistant, but not waterproof.

If the Ax60+ sensor unit gets soaked, it can sometimes cause it to go into alarm in error.

Powering down the sensor unit and giving it time to dry out should solve the problem and the Ax60+ should return to normal operation.

What height should my sensors be installed?
  • CO₂ – 12-18″ above ground
  • Cultivate  – 12-18″ above ground
  • SAFE-OX – Average Head height
  • O₂ – Average Head height
  • CO – Average Head height
How to silence and acknowledge latched alarms on the Analox Ax60+

The Ax60+ systems have latched alarms as standard however does have the option of disabling alarm latching. It is our hope that the following information will help to avoid confusion when silencing and acknowledging alarms.

Latched Alarms

These will not clear unless acknowledged. For example, if the target gas level increases beyond the alarm threshold, the device will go into alarm and remain like this until the alarm has been acknowledged.

Non-Latched Alarms

These will clear without acknowledgement. For example, if the target gas level increases beyond the alarm threshold, the unit will go into alarm but once the gas returns to safe limits the alarm automatically clears.

Both Latched and Non-Latched alarms can be muted with a single press of the Accept/Test button, but in order to acknowledge a latched alarm the user must first cycle through the display until an ‘*’ is seen, then press and hold the Accept/Test button for 3 seconds (resulting in the ‘*’ changing into a ‘>’.

The video below shows an example of muting and acknowledging a latched alarm:

 

How do I fix a loose RJ45 connector on my Ax60+?

What should I do if my Ax60+ has a FLT16 on the display?

All relevant Ax60+ document information can be found on the Ax60+ product page.

There are two potential reasons for the Ax60+ to show a FLT16 code on the screen, these are:

  1. The sensor unit requires zero and span calibration adjustment (see the service manual for guidance on carrying out this process)
  2. The CO₂ sensor alarm directions have been incorrectly set.

If your unit is displaying a Flt16, first attempt a zero and span calibration adjustment then power cycle the device. If this is not possible or the issue is not resolved, check to see if it has been correctly configured as detailed below.

The Ax60+ is primarily a breathing air device, as such, the CO₂ sensor unit alarms are restricted to only trigger in the ‘high-going’ direction. If a user changes the alarm ‘going’ to ‘low-going’ a FLT16 will ensue.

To clear this fault the following steps should be followed (be aware that only a short period of non-alarm time is available so you’ll need to be quick!!):

  1. Power cycle the Ax60+ (turn it off and on again)
  2. Wait for the sensors display on the screen
  3. Press the Accept/Test button briefly to silence any alarms. Press and hold the top two buttons (Cancel and cycle) until the top level menu is reached.
  4. Press the cycle button six times (until sensor config is shown)
  5. Press the Accept/Test button. Press the cycle button twice (Alarm direction is shown)Press the Accept/Test button. Select the sensor that had the alarm ‘going’
  6. changed (press cycle to move the arrow selector over until the relevant sensor is identified and press the Accept/. test button)
  7. Select the alarm that had the alarm ‘going’ changed
  8. (press cycle to move the arrow selector over until the relevant alarm is identified and press the Accept/test button)
  9. Change the alarm direction to ‘up’ (press the cycle button until the arrow selector points at the up arrow and press the Accept/test button)
  10. Press the cancel button five times until the sensors display on screen
  11. Acknowledge any faults displayed
  12. Power cycle the unit

Is the Ax60+ backwards compatible with the Ax60?

There is some backwards compatibility:

  • Ax60+ CO₂ sensor units CAN be used with both Ax60 and Ax60+ central displays.
  • Ax60 sensor units CAN ONLY be used with Ax60 central displays.
  • Oxygen sensors CAN ONLY be used on the Ax60+ central displays.

See table below for reference:

 

Ax60 Central Display

Ax60+ Central Display

Ax60 CO₂ Sensor

✔

✗

Ax60+ CO₂ Sensor

✔

✔

Ax60+ O₂ Sensor

✗

✔